Original ResearchNormative Data for pH of Exhaled Breath Condensate
Section snippets
Subjects
Self-described healthy volunteers were recruited from the area surrounding the University of Virginia in the United States. This is a mixed community of urban, suburban, and rural families. To ensure that the population would serve as the most useful control group, subjects were excluded from the study if they affirmed on screening that they smoked tobacco or had any of the following symptoms or diseases: chronic cough, recurrent or chronic wheeze, COPD, emphysema, chronic bronchitis, cystic
Results
Four hundred twelve subjects were enrolled and provided a single EBC sample for this study. The subject population is described in Table 2. Of the 412 subjects enrolled, 404 provided sample volumes sufficient for EBC pH assay, a success rate > 98%.
The median EBC pH of this healthy population was 8.0 with interquartile (25 to 75%) range of 7.8 to 8.1 (Fig 1). The maximum EBC pH was 8.4, and the minimum pH was 4.5. The values were the same for female patients (median, 8.0; interquartile range,
Discussion
In this study, we have demonstrated that in healthy subjects, the median, gas-standardized EBC pH is 8.0 with an interquartile range of 7.8 to 8.1. The data are tightly clustered in a mildly nonparametric bell curve. This normal range provides additional context to the published data revealing prominently lower values in patients with multiple respiratory disease states (with EBC pH often in the pH 4 to 6 range).123578121314 Our data present a range that is slightly higher than that from
References (21)
- et al.
Breath condensate in children with cystic fibrosis and asthma: a new noninvasive marker of airway inflammation?
Chest
(2004) - et al.
Exhaled breath condensate acidification in acute lung injury
Respir Med
(2003) - et al.
Acid stress in the pathology of asthma
J Allergy Clin Immunol
(2004) - et al.
Inadvertent bronchial pH monitoring [letter]
Gastrointest Endosc
(1987) - et al.
Twenty-four-hour tracheal pH monitoring: a simple and non-hazardous investigation
Respir Med
(1994) Gastroesophageal reflux as an asthma trigger: acid stress
Chest
(2004)- et al.
Endogenous airway acidification: implications for asthma pathophysiology
Am J Respir Crit Care Med
(2000) - et al.
pH in expired breath condensate of patients with inflammatory airway diseases
Am J Respir Crit Care Med
(2002) - et al.
Acid-base equilibrium in exhaled breath condensate of allergic asthmatic children
Allergy
(2005) - et al.
Exhaled breath condensate pH and exhaled nitric oxide in allergic asthma and in cystic fibrosis
Thorax
(2005)
Cited by (120)
Exhaled breath condensate analysis in horses: A scoping review
2024, Research in Veterinary ScienceBiomarkers for the adverse effects on respiratory system health associated with atmospheric particulate matter exposure
2022, Journal of Hazardous MaterialsReduced Exhaled Breath Condensate pH and Severity of Allergic Sensitization Predict School Age Asthma
2021, Journal of Allergy and Clinical Immunology: In PracticeNatural protection of ocular surface from viral infections – A hypothesis
2020, Medical HypothesesExhaled Breath Condensate: An Update
2018, Immunology and Allergy Clinics of North AmericaAssessment of metals in detection of TB biomarkers: Novel computational approach
2015, Materials Chemistry and Physics
Reproduction of this article is prohibited without written permission from the American College of Chest Physicians (www.chestjournal.org/misc/reprints.shtml).
Dr. Hunt is a co-founder of Respiratory Research, Inc, the company that manufactures the EBC collectors used in this study. This company and the University of Virginia own patents relating to collection of and pH measurement in EBC.