Compton-scatter tissue densitometry: calculation of single and multiple scatter photon fluences

and

Published under licence by IOP Publishing Ltd
, , Citation J J Battista and M J Bronskill 1978 Phys. Med. Biol. 23 1 DOI 10.1088/0031-9155/23/1/001

0031-9155/23/1/1

Abstract

The accurate measurement of in vivo electron densities by the Compton-scatter method is limited by attenuations and multiple scattering in the patient. Using analytic and Monte Carlo calculation methods, the authors have modelled the Clarke tissue density scanner for incident monoenergetic photon energies from 300 to 2000 eV and for mean scattering angles of 30 to 130.

Export citation and abstract BibTeX RIS