Sol–gel-based amperometric glucose biosensor incorporating an osmium redox polymer as mediator

Abstract

A novel amperometric biosensor for the determination of glucose was constructed by first immobilizing glucose oxidase and an osmium redox polymer, [Os(bpy)2(PVP)10Cl]Cl, on the surface of a glassy-carbon electrode, followed by coating with a sol–gel film derived from methyltriethoxysilane. The redox potential of this electrode was found to be +247 mV versus Ag/AgCl, and glucose could therefore be determined amperometrically at +400 mV versus Ag/AgCl. The concentration range of linear response, slope of linear response and LOD were 2.0–31 mmol l–1, 44.5 nA mmol–1 l and 0.5 mmol. l–1, respectively. Although L-ascorbate was electrooxidizable at this potential, uric acid and paracetamol were both found not to interfere.

Article information

Article type
Paper

Anal. Commun., 1996,33, 271-273

Sol–gel-based amperometric glucose biosensor incorporating an osmium redox polymer as mediator

T. Park, E. I. Iwuoha, M. R. Smyth and B. D. MacCraith, Anal. Commun., 1996, 33, 271 DOI: 10.1039/AC9963300271

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Spotlight

Advertisements