Abstract
The push-out test of three types of biomaterials: carbon fibre-reinforced carbon (CFRC), hydroxyapatite (HA), and surgical steel (SS) implanted into rabbits’ femurs was carried out. Hydroxyapatite was used as a positive control (good fixation expected in bone) and surgical steel was a negative one (potentially no fixation in bone). Regeneration of bone in contact with all implants was found three months after implantation. The shear strength between CFRC implants and bone was lower than with the HA implants and higher than the shear strength between the surgical steel and bone. Compressive strength of CFRC implants removed after the observation period was significantly lower than the compressive strength of non-implanted samples. It is concluded that the mechanical bonding between the CFRC implants and host tissues exists 3 months after intrabone implantation and is accompanied by a decrease of the strength of implants.
Similar content being viewed by others
References
P. CHRISTEL and L. CLAES, in “High performance biomaterials —a comprehensive guide to medical and pharmaceutical applications”, edited by M. Szycher (Technomic Publishing Company, Lancaster, 1991) p. 499.
D. ADAMS, D. F. WILLIAMS and J. HILL, J. Biomed. Mater. Res. 12 (1978) 35.
L. PODOSHIN, R. H. NODAR, G. B. HUGHES, T. BAUER, J. D. HAYES, M. FRADIS, J. BOSS and L. RAMSEY, Amer. J. Otol. 9 (1988) 366.
Y. SHONO, P. C. Mc AFEE, B. W. CUNNINGHAM and J. W. BRANTIGAN, J. Bone Joint. Surg. Amer. 75 (1993) 1674.
C. L. TIAN, V. J. HETHERINGTON and S. REED, J. Foot Ankle 32(5) (1993) 490.
F. VALLANA, E. PASQIUNO, S. RINALDI, M. GALLONI, A. M. GATTI, F. MODICA and A. BENECH, Ceramics Int. 19 (1993) 169.
K. HAYASHI, T. INADOME, H. TSUMURA, Y. NAKASHIMA and Y. SUKOIOKA, Biomaterials 15 (1994) 1187.
T. INADOME, K. HAYASHI, Y. NAKASHIMA, H. TSUMURA and Y. SUGIOKA, J. Biomed. Mater. Res. 29 (1995) 19.
H. KIEFER, L. CLAES, C. BURRI and K. KUGLMEIER, in “Biological and biomechanical performance of biomaterials”, edited by P. Christel, A. Meunier and A. J. C. Lee (Elsevier Science Publishers B.V., Amsterdam, 1986) p. 471.
K. SØBALLE, Acta Orthop. Scand. 64(Suppl 255) (1993) 1.
J. CHŁOPEK, S. BŁAŻEWICZ, Carbon 29 (1991) 127.
M. LEWANDOWSKA-SZUMIEŁ, J. KOMENDER, D. KUDELSKA and M. MAZUR, in “Animal alternatives, welfare and ethics”, edited by L. F. M. van Zutphen and M. Balls (Elsevier Science Publishers B.V., Amsterdam, 1997) in press.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
LEWANDOWSKA-SZUMIEŁ , M., KOMENDER , J., GORECKI , A. et al. Fixation of carbon fibre-reinforced carbon composite implanted into bone. Journal of Materials Science: Materials in Medicine 8, 485–488 (1997). https://doi.org/10.1023/A:1018526226382
Issue Date:
DOI: https://doi.org/10.1023/A:1018526226382