Elsevier

Biomaterials

Volume 10, Issue 8, October 1989, Pages 511-516
Biomaterials

Paper
Intraocular PMMA lenses modified with surface-immobilized heparin: evaluation of biocompatibility in vitro and in vivo

https://doi.org/10.1016/0142-9612(89)90055-0Get rights and content

Abstract

Intraocular lenses (IOL) were surface modified with covalently linked heparin. The surface-bound heparin could not be removed by incubation in solutions known to be effective in breaking non-covalent bonds, nor by incubation in a solution of proteinase K and only to a limited extent by incubation with heparinase. In vitro studies demonstrated improved biocompatibility by the heparin surface-modified lens with respect to outgrowth of fibroblasts and macrophages, activation of granulocytes and adhesion of platelets. These results were subsequently verified in vivo in terms of less inflammatory cells on the lens surface and fewer incidences of synechiae after 3 and 6 wk IOL implantation in the rabbit eye.

Cited by (104)

  • Long-term In Vivo Stability of Posterior Chamber Phakic Intraocular Lens: Properties and Light Transmission Characteristics of Explants

    2020, American Journal of Ophthalmology
    Citation Excerpt :

    These are well-known materials to which the immune system of the living host hardly reacts. In the 1980s, it was determined that coating the surface of a lens with another substance reduced the host's response to apIOLs.19 Furthermore, a soft contact lens made by copolymerizing methacrylate with collagen was commercially available.

  • Synthetic Biomaterial for Regenerative Medicine Applications

    2017, Kidney Transplantation, Bioengineering, and Regeneration: Kidney Transplantation in the Regenerative Medicine Era
  • Laser-triggered intraocular implant to induce photodynamic therapy for posterior capsule opacification prevention

    2016, International Journal of Pharmaceutics
    Citation Excerpt :

    Co., Ltd (Changchun, People’s Republic of China). Preparation method of ICG-IOL was similar to that of modifying IOLs with heparin (Larm et al., 1983; Larsson et al., 1989). Briefly, IOLs were first sonicated in a bath of surfactant solution for 15 min, and then in double distilled water for another 15 min.

  • Effect of heparin coating of a foldable intraocular lens on inflammation and capsular bag performance after cataract surgery

    2013, Journal of Cataract and Refractive Surgery
    Citation Excerpt :

    In this study, heparin-coated foldable acrylic IOLs did not show any benefit over the same IOL without heparin coating in terms of postoperative inflammation in high-risk patients. This finding is contrary to results in older studies in which HSM PMMA IOLs were used and found to effectively reduce postoperative inflammation2,3,6,23–28 because of the anticoagulant and antiinflammatory effects of heparin.20–22 But, when HSM PMMA IOLs were compared with uncoated foldable acrylic IOLs, no significant differences in flare were found.36–38

  • Diamond like carbon films as a protective surface on PMMA for biomedical applications

    2010, Surface and Coatings Technology
    Citation Excerpt :

    The life span of PMMA prostheses will also increase by the DLC coating. Extensive research has been carried out on the biocompatible study of PMMA and DLC coatings, separately [2–5,12,13,18–20]. There are very few reports on the detailed study of biocompatibility of DLC coated PMMA [8,9].