The new grasper-integrated single use flexible cystoscope for double J stent removal: evaluation of image quality, flow and flexibility

  • Original Article
  • Published:
World Journal of Urology Aims and scope Submit manuscript

Abstract

Background

A new single use digital flexible cystoscope (FC) Isiris α from Coloplast® with an incorporated grasper has been developed to perform double J stent removal. There is a lack of data regarding the comparison of image quality, flexibility and flow between classic cystoscopes and the new Isiris α.

Materials and methods

Five different FC were used to compare the image quality, the field of view, the loss of flow and the deflection loss. Two standardized grids, three stones of different composition and a ruler’s image were filmed in four standardized different scenarios. These videos were shown to thirty subjects that had to evaluate them. Water outflow was measured in ml/sec in all devices with and without the grasper inside, instruments tip deflection was measured using a software.

Results

In the subjective analysis of the image quality Isiris α was the second FC best scored. At 3 cm of distance, the field view of Isiris α was the narrowest. Comparing the water flow in the different FCs, we observed a water flow decrease in all cystoscopes when the grasper was loaded in the working channel. Isiris α deflection and flow increase when the grasper is activated.

Conclusion

In terms of quality of vision and water flow, the FC Isiris α is comparable to the other digital FC tested. Field of view is narrower. The results displayed a valid alternative to the standard procedure for DJ removal.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Quayle SS, Ames CD, Lieber D, Yan Y, Landman J (2005) Comparison of optical resolution withdigital and standard fiberoptic cystoscopes in an in vitro model. Urology 66(3):489–493

    Article  PubMed  Google Scholar 

  2. Abdelshehid C, Ahlering MT, Chou D, Park HK, Basillote J, Lee D, Kim I, Eichel L, Protsenko D, Wong B, McDougall E, Clayman RV (2005) Comparison of flexible ureteroscopes: deflection, irrigant flow and optical characteristics. J Urol 173(6):2017–2021

    Article  PubMed  Google Scholar 

  3. Lusch A, Liss MA, Greene P, Abdelshehid C, Menhadji A, Bucur P, Alipanah R, McDougall E, Landman J (2013) Comparison of optics and performance of a distal sensor high definition cystoscope, a distal sensor standard definition cystoscope, and a fiberoptic cystoscope. Urology 82(6):1226–1230

    Article  PubMed  Google Scholar 

  4. Isiris α™ Coloplast Instructions for Use Manual

  5. USAF 1951 From Web site of Edmund Optics, Barrington, NJ

  6. Color Checker. From Web site of Edmund Optics, Barrington, NJ

  7. Clemens JQ, Dowling R, Foley F, Goldman HB, Gonzalez CM, Tessier C, Wasner MA, Young E Joint AUA/SUNA White Paper on Reprocessing of Flexible Cystoscopes. From the American Urological Association Education and Research, Inc., the Society of Urologic Nurses and Associates, the Department of Urology, University of Michigan Medical Center, Ann Arbor, Michigan (JQC), Urology Associates of North Texas, Arlington, Texas (RD), Saint Vincent Hospital, Worcester (FF), and Olympus ACMI, Southborough (EY), Massachusetts, the Glickman Urological and Kidney Institute, Cleveland Clinic, Cleveland, Ohio (HBG), Department of Urology, Northwestern University Feinberg School of Medicine (CMG) and Center for Urologic Health, Advocate Illinois Masonic Medical Center (MAW), Chicago, Illinois, and Manchester Urology Associates, Manchester, New Hampshire (CT)

  8. Bretthauer M, Jørgensen A, Kristiansen BE, Hofstad B, Hoff G (2003) Quality control in colorectal cancer screening: systematic microbiological investigation of endoscopes used in the NORCCAP (Norwegian Colorectal Cancer Prevention) trial. BMC Gastroenterol 3:15–18

    Article  PubMed  PubMed Central  Google Scholar 

  9. Fraser V, O’Rourke S, Jones M, Murray P, Clouse RE, Klasner J et al (1992) Gastrointestinal endoscope disinfection: a prospective randomized trial comparing automated and manual disinfection [Abstract]. Gastrointest Endosc 38:277. http://www.ncbi.nlm.nih.gov/pubmed/8354869

  10. Okhunov Z, Hruby GW, Mirabile G, Marruffo F, Lehman DS, Benson MC, Gupta M, Landman J (2009) Prospective comparison of flexible fiberoptic and digital cystoscopes. Urology 74(2):427–430. doi:10.1016/j.urology.2009.01.007

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Dr. Esteban Emiliani is supported by the EUSP scholarship. No financing has been received to perform this study.

Author’s contribution

Talso: Protocol/project development; Data collection and management; Data analysis;Manuscript writing and Editing. Emiliani: Protocol/project development; Data collection and management, Manuscript editing. Baghdadi: Protocol/project development; Data collection and management. Orosa: Data collection and management; Data analysis. Servian: Protocol/project development; Data collection and management; Data analysis. Barreiro: Protocol/project development; Data collection and management. Proietti: Protocol development. Traxer: Protocol/project development; Data collection and management; Data analysis; Manuscript writing/editing.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. Talso.

Ethics declarations

Ethical standard

The material in the manuscript has been acquired according to modern ethical standards and has been approved by the legally appropriate ethical committees.

Conflict of interest

Professor Traxer is a consultant for the following companies: Coloplast, Rocamed, Olympus, Boston Scientific, Lumenis, BioHealth.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Talso, M., Emiliani, E., Baghdadi, M. et al. The new grasper-integrated single use flexible cystoscope for double J stent removal: evaluation of image quality, flow and flexibility. World J Urol 35, 1277–1283 (2017). https://doi.org/10.1007/s00345-016-1987-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00345-016-1987-z

Keywords

Navigation