Abstract.
An analytical study on the use of graphite-epoxy composite (GEC) electrodes for differential pulse anodic stripping voltammetry (DPASV) of heavy metals is presented. This study is accompanied by microscopic observations of the electrode surface before and after the stripping step in comparison to glassy carbon electrode. GEC electrodes show much better accumulation properties and consequently acceptable behaviour which makes them suitable as working electrodes in the DPASV of heavy metals. Lead determination in real water samples in a batch system as well as some preliminary results in a flow-through system are presented. The detection limits in batch measurements were 100 ppb for Cd, 10 pb for lead and 50 ppb for copper. The detection limit for lead in the flow-through system was similar to that in the batch. The results obtained show that these low cost and easy to prepare materials can be of interest in future research concerning stripping techniques of heavy metals and other analytes.
Similar content being viewed by others
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Carrégalo, S., Merkoçi, A. & Alegret, S. Application of Graphite-Epoxy Composite Electrodes in Differential Pulse Anodic Stripping Voltammetry of Heavy Metals. Microchim. Acta 147, 245–251 (2004). https://doi.org/10.1007/s00604-004-0229-z