Abstract
The magnitude, mismatch and temporal variations of the electrode-gel-skin interface impedance can cause problems in electrical impedance tomography (EIT) measurement. It is shown that at the high frequencies generally encountered in EIT the capacitive properties of the electrode interface, and especially those of the skin, are of primary importance. A wide range of techniques are reviewed that could possibly be used to minimise these problems. These techniques include the use of skin preparation, penetration enhancers, temperature and electrical impulses. Although several of these techniques appear very attractive, they are not without serious potential drawbacks. A combination of some of these techniques may well hold the key to success.
Similar content being viewed by others
References
Baisch, F. J., Hahn, G., Šipinková, I., Beer, M., andHellige, G. (1995): ‘Comparison of electrode belts with ‘spot’ electrodes in electrical impedance tomography’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Techno. Bio. Med.,16, (2), pp. 119–125
Barber, D. C., Brown, B. H., andFreeston, I. L. (1983): ‘Imaging spatial distributions of resistivity using applied potential tomography’,Electron Lett.,19, pp. 933–935
Bard, A. J. andFaulkner, L. R. (1980): ‘Electrochemical methods’ (John Wiley)
Boone, K., andHolder, D. S. (1995a): ‘Assessment of noise and drift artefacts in electrical impedance tomography measurements using the Sheffield Mark 1 system’, Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Technol. Biol. Med.,16, (2), pp. 49–60
Boone, K., andHolder, D. S. (1995b): ‘A model of the effect of variations in contact and skin impedance on electrical impedance tomography measurement artefacts’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Techno., Bio. Med.,16, (2), pp. 61–70
Brown, B. H., Barber, D. C., Wang, W., Lu, L., Leatherhead, A. D., Smallwood, R. H., Hampshire, A., Mackay, R., andHatzigalamis, K. (1994): ‘Multi-frequency imaging and modelling of respiratory related electrical impedance changes’,Physiol. Meas. 15, supplement A, A1-A12
Brown, L., andLanger, R. (1988): ‘Transdermal delivery of drugs’,Ann. Rev. Med.,39, pp. 221–229
Campbell, S. D., Kraning, K. K., Schibli, E. G., andMommi, S. T. (1977): ‘Hydration characteristics and electrical resistivity of stratum corneum using a non-invasive four-point microelectrode method’,J. Invest. Dermatol.,69, pp. 290–295
Carim, H. M. (1988): ‘Bioelectrodes’in Webster, J. G. (Ed.): ‘Encyclopedia of medical devices and instrumentation’ (John Wiley and Sons, New York) pp. 195–226
Chien, Y. M. (1982): ‘Transdermal controlled-release drug administration’in Swarbrick, J. (Ed.): ‘Novel drug delivery systems’ (Marcel Dekker Inc., New York) p. 149
Chien, Y. W. (1987): ‘Development of transdermal drug delivery systems’,Drug Devel. Ind. Pharmacy,13,(4&5), pp. 589–651
Chien, Y. W., andLee, C-S. (1987): ‘Transdermal drug delivery system with enhanced skin permeability’,Am. Chem. Soc., Symp. Series,34E pp. 281–300
Cullander, C. (1992): ‘What are the pathways of iontopheritic current flow through mammalian skin?’,Adv. Drug Delivery Rev.,9, pp. 119–135
Cullander, C., andGuy, R. H. (1992): ‘Visualization of iontophoretic pathways with confocal microscopy and the vibrating probe electrode’,Solid State Ionics, pp. 53–56, 197–206
De Levie, R. (1965): ‘The influence of surface roughness of solid electrodes on electrochemical measurements’,Electrochim. Acta,10 pp. 113–130
Edelberg, R. (1971): ‘Electrical properties of the skin’in Elden, H. R. (Ed.) ‘A treatise of the skin’ (John Wiley & Sons)
Foley D., Corish J., andCorrigan O. I. (1992): ‘Iontophoretic delivery of drugs through membranes including human stratum corneum’,Solid State Ionics, pp. 53–56, 184–196
Fuks, L. F., Cheney, M., Isaacson, D., Gisser, D. G., andNewell, J. C. (1991): ‘Detection and imaging of electric conductivity and permittivity at low frequency’,IEEE Trans.,BME-38, pp. 1106–1110
Gatzke, R. D. (1974): ‘The electrode: a measurement systems viewpoint’in Miller, H. A., andHarrison, D. C.: ‘Biomedical electrode technology’ (Academic Press, New York) pp. 99–116
Gersing E., Schafer, M., andOsypka, M. (1995): ‘The appearance of positive phase angles in impedance measurements on extended biological objects’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Technol. Biol. Med.,16, (2), pp. 71–76
Gerstner H., andGerbstadt, H. (1949): ‘Der wechselstromwider-stand der menschlichen Haut’,Arch. f. d. ges. Physiol.,252 p. 111
Goble, J., andIsaacson, D. (1989): ‘Optimal current patterns for three-dimensional electric current computed tomography’, Proc. Ann. Int. Conf. of IEEE Engineering in Medicine and Biology Society,11, pp. 463–464
Gondran, C., Siebert, E., Yacoub, S., andNovakov, E. (1995): ‘Dry electrode based on nasicon ceramic for surface biopotential measurements’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UKInnov. Technol. Biol. Med.,16, (2), pp. 113–118
Grimnes, S. (1983a): ‘Dielectric breakdown of human skin in vivo’,Med. Biol. Eng. Comput.,21 pp. 379–381
Grimnes, S. (1983b): ‘Skin impedance and electro-osmosis in the human epidermis’Med. Biol. Eng. Comput.,21, pp. 739–749
Grimnes S. (1983c): ‘Impedance measurement of individual skin surface electrodes’,Med. Biol. Eng. Comput.,21, pp. 750–755
Holder, D. S. (1995): ‘Design and electrical characteristics of a multi-level electrode array for electrical impedance tomography of the female breast’. Proc. Concerted Action on Impedance Tomographay (CAIT). Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Technol. Biol. Med.,16, (2), pp. 143–150
Isaacson, D. (1986): ‘Distinguishability of conductivities by electric current computed tomography’,IEEE Trans.,M-5,(2), pp. 92–95
Jossinet, J. (1988): ‘A hardware design for imaging the electrical impedance of the breast’,Clin. Phys. Physiol. Meas.,9, supplement A, pp. 25–8
Jossinet, J., McAdams, E., McLaughlin, J., andJarry, R. (1995): ‘The interface impedance of metal electrodes for electrical impedance tomography’. Proc. Concerted Action on Impedance Tomography (CAIT). Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Technol. Biol. Med.,16, (2), pp. 103–112
Jossinet, J., andTrillaud, C. (1992): ‘Technical improvement of a dual frequency EIT tomograph for bioelectrical characterisation’. Proc. 14th Ann. Int. Conf. of IEEE EMBS Society, 1712
Kalia, Y. K., andGuy, R. H. (1995.): ‘The electrical characteristics of human skin in vivo’.Pharm. Res.,12 (11), pp. 1605–1613
Kasting, G. B. (1992): ‘Theoretical models for iontophoretic delivery’,Adv. Drug Delivery Res.,9, pp. 177–199
Kelly, H. W. (1985): ‘Controlled-release transdermal drug delivery’,Cutis,35, pp. 204–207
Klingman, A. M. (1984): ‘Skin permeability: dermatologic aspects of transdermal drug delivery’,Am. Heart.,108, (1), pp. 200–207
Knepp, V. M., Hadgraft, J., andGuy, R. H. (1987): ‘Transdermal drug delivery: problems and possibilities’,CRC Crit. Rev. Therapeutic Drug Carrier Syst.,4 (1), pp. 13–37
Kontturi, K., Murtomäki, L., Hirvonen, J., Paronen, P., andUrtti, A. (1993): ‘Electrochemical characterization of human skin by impedance spectroscopy: The effect of penetration enhancers’Pharm. Res.,10, (3), pp. 381–385
Krenzke, G., Dusterhoft T., andSchwanke, R. (1985): ‘Vorrichtung zur Albeitung von Herzpol’. German Patent DD 225 333 Al
Lawler, J. C., Davis M. J., andGriffith, E. C. (1960): ‘Electrical characteristics of the skin’, J. Invest. Dermatol., pp. 301–308
McAdams, E. T. (1989a): ‘Effect of surface topography on the electrode-electrolyte interface impedance (i) the high frequency, small signal interface impedance—a review’,Surface Topography,2, pp. 107–122
McAdams, E T. (1989b): ‘Effect of surface topography on the electrode-electrolyte interface impedance (ii) the low frequency, small signal, interface impedance— a review’,Surface Topography,2, pp. 223–232
McAdams, E. T. (1990): ‘Surface biomedical electrode technology’,Int. Med. Device Diagnostic Ind., pp. 44–48
McAdams, E. T., andJossinet, J. (1990): ‘Hydrogel electrodes in bio-signal recording’. Proc. 12th Ann. Int. Conf. of IEEE, Engineering in Medicine and Biology Society, Philadelphia, USA, pp. 1490–1491
McAdams, E. T., andJossinet, J. (1991a): ‘The importance of electrode-skin impedance in high resolution electrocardiography’,Automedica,13, pp. 187–208
McAdams, E. T., andJossinet, J. (1991b): ‘DC nonlinearity of the elcectrode-electrolyte interface impedance’,Innov. Technol. Biol. Med.,12 (3), pp. 329–343
McAdams, E. T., andJossinet, J. (1992): ‘A physical interpretation of Schwan's limit current of linearity’,Ann. Biomed. Eng.,20 (3), pp. 307–319
McAdams, E. T., andJossinet, J. (1994): ‘A physical interpretation of Schwan's limit voltage of linearity’Med. Biol. Eng. Comput.,32, (3), pp. 126–130
McAdams, E. T., andJossinet, J. (1995): ‘Tissue impedance: a historical overview’,Physiol. Meas.,16, A1-A13
McAdams, E. T., McLaughlin, J. A., andHolder, D. S. (1992a): ‘Neurosensors: A review of some fundamental electrode parameters’. Satellite Symp. on Neuroscience and Technology, 14th Ann. Conf. of the IEEE Engineering in Medicine and Biology Society, Lyon, France, pp. 226–234
McAdams, E. T., Henry, P., andAnderson, J. McC. (1992b): ‘Optimal electrolytic chloriding of silver ink electrodes for use in electrical impedance tomography’,Clin. Phys. Physiol. Meas.,13, supplement A, pp. 19–23
McAdams, E. T., McLaughlin, J. A., Anderson J. McC. (1994a): ‘Multi-electrode systems for electrical impedance tomography’,Physiol. Meas.,15, A101-A106
McAdams, E. T., McLaughlin, J. A., Brown, B. N., andMcArdle F. (1994b): ‘The NIBEC EIT harness’in Holder, D. (Ed.) ‘Clinical and physiological applications of electrical impedance tomography’ (UCL Press, London, 1993) Chap. 8, pp. 85–92
McAdams, E. T., Lackermeier, A., andJossinet, J. (1994c): ‘AC impedance of the hydrogel-skin interface’ 16th Ann. Int. Conf. of IEEE Engineering in Medicine and Biology Society, Baltimore. USA, pp. 870–871
Newell, J. C. (1995): ‘Influence of electrode size on in-vivo impedance images’. Concerted Action on Impedance Tomography (CAIT). Workshop on The Reproducibility of in vivo Electrical Impedance Tomography Images, 12–14 May, Keele, UK
Nolan, L. M. A., Corish, J., andCorrigan, O. I. (1993): ‘Electrical properties of human stratum corneum and transdermal drug transport’,J. Chem. Soc. Faraday Trans.,89, (15), pp 2839–2845
Oh, S. Y., Leung, L., Bommannan, D., Guy R. H., andPotts, R. O. (1993): ‘Effect of current, ionic strength and temperature on the electrical properties of skin’,J. Controlled Release,27, pp. 115–125
Ollmar, S. (1995): ‘Factors inflencing the electrical properties of the skin’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Technol. Biol. Med.,16 (2), pp. 136–142
Paulson K., Breckon, W., andPidcock, M. (1992): ‘Electrode modelling in electrical impedance tomography’,SIAM J. Appl. Math.,52, pp 1012–1022
Phipps, J. B., andGyory, J. R. (1992): ‘Transdermal ion migration’,Adv. Drug Delivery Rev.,9 pp. 137–176
Pikal, M. J. (1992): ‘The role of electroosmotic flow in transdermal iontophoresis’,,9, pp. 201–237
Prausnitz, M. R., Bose, V. G., L&S:Anger, R., andWeaver, J. C. (1993): ‘Electroporation of mammalian skin: A mechanism to enhance transdermal drug delivery’,Proc. Nat. Acad. Sci.,90, pp. 10504–10508
Reilly, J. P. (1992): ‘Electrical stimulation and electropathology’ (Cambridge University Press)
Rein, H. (1924): ‘Eperimentelle Studien über Elecktroendosmose an überlebender menschlicher’,Haut. Z. Biol.,81 pp. 125–140
Rigaud, B., Hamzaoui, L., Granie, M., Chauveau N., Martinez E., andMorucci, J. P. (1995): ‘Electrode-electrolyte interface impedance estimation beyond 10 Hz for different electrolytes and electrodes’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UKInnov. Technol. Biol. Med.,16, (2), pp. 87–94
Risacher, F., Jossinet, J., McAdams, E. T., Mann, Y., andSchmitt, M. (1993): ‘Impedance plethysmography for the evaluation of pulse wave velocity in limbs’,Med. Biol. Engl. Comput.,31, (3), pp. 318–322
Risacher, F. (1995): ‘Etude de la propagation de l'onde de pouls par pléthsmographie d'impedance électrique’. PhD Thesis, l'Universitié Claude Bernard, Lyon, France
Riu, P. J., Rosell, J., Lozano, A., andPallas-Areny, R. (1995): ‘Multi-frequency static imaging in electrical impedance tomography: Part 1 Instrumentation requirements’,Med. Biol. Eng. Comput.',33, pp. 784–792
Rosell, J., Colominas, J., Riu, P., Palas, R., andWebster J. G. (1988): ‘Skin impedance from 1 Hz to 1 MHz’,IEEE Trans.,BME-35, pp. 649–651
Rosendal, T. (1943): ‘Studies on the conducting properties of the human skin to direct current’Acta. Physiol. Scand.,5, pp. 130–151
Salter, D. C. (1981): ‘A study of some electrical properties of normal and pathological skin in vivo’. D Phil Thesis, University of Oxford
Schmitt, O. H., andAlmasi, J. J. (1970): ‘Electrode impedance and voltage offset as they affect efficacy and accuracy of VCG and ECG measurements’. Proc. X1th Int. Vectorcardiography Symp., New York, pp. 245–253
Smith, R. W. M. (1990): ‘Design of a real time impedance imaging system for medical applications’. PhD Dissertation, University of Sheffield, pp. 45–49
Taktak, A., Record, P. M., Gadd, R., andRolfe, P. (1995a): ‘Interface impedance measurement of paediatric electrodes’ Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography, Belfast, UK,Innov. Techno. Biol. Med.,16 (2), pp. 126–135
Taktak, A., Record, P.M., Gadd R., andRolfe, P. (1995b): ‘Practical factors in neonatal lung imaging using electrical impedance tomography’,Med. Biol. Eng. Comput.,33, pp. 202–205
Wang, M. (1994): ‘Electrical impedance tomography on conducting walled process vessels’, PhD Dissertations, UNIST, UK
Weaver J. C. (1993): ‘Electroporation—a general phenomenon for manipulating cells and tissues’,J. Cell. Biochem.,51, (4), pp. 426–435
Wiechers J. (1992): ‘Use of chemical penetration enhancers in transdermal drug delivery—possibilities and difficulties’Acta Pharm. Nord. 4, (2), pp. 113–128
Yamamoto, T., andYamamoto, Y. (1976): ‘Electrical properties of the epidermal stratum corneum’,Med. Biol. Eng., pp. 151–158
Yamamoto, Y., andYamamoto, T. (1978): ‘Dispersion and correlation of the parameters for skin impedance’,Med. Biol. Eng. Comput.,16, pp. 592–594
Yamamoto, T., andYamamoto, Y. (1981): ‘Non-linear electrical properties of skin in the low frequency range’,Eng. Comput.,19, pp. 302–310
Yamamoto, T., andYamamoto, Y., andYoshida A. (1986): ‘Formative mechanisms of current concentration and breakdown phenomena dependent on direct current flow through the skin by a dry electrode’,IEEE Trans,BME-33, (4), pp. 386–404
Yorkey, T. J., Webster, J. G., andTompkins, W. J. (1985): ‘Errors caused by contact impedance imaging’. Proc. 7th Ann. Int. Conf. of IEEE Engineering in medicine and Biology Society, pp. 632–637
Zhou, D. M., McAdams. E., Lackermeier, A., andJones J. G. (1994): ‘AC impedance of Ag-AgCl reference electrodes for use in disposable biosensors’. Proc. 16th Ann. Int. Conf. of IEEE Engineering in Medicine and Biology Society, Baltimore, USA, pp. 832–833
Zhou, D. M., Jones J. G., McAdams, E., Lackermeier, A. (1995): ‘Study of the electrode-electrolyte interface of screen printed Ag-AgCl electrodes’. Proc. Concerted Action on Impedance Tomography (CAIT) Workshop on The Electrode-Skin Interface in Electrical Impedance Tomography’, Belfast, UK,Innov. Techno. Biol. Med.,16, (2), pp. 77–86
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
McAdams, E.T., Jossinet, J., Lackermeier, A. et al. Factors affecting electrode-gel-skin interface impedance in electrical impedance tomography. Med. Biol. Eng. Comput. 34, 397–408 (1996). https://doi.org/10.1007/BF02523842
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02523842