Abstract
The established coincidence method of determining125I activities has been adapted for use with a multisample gamma counter. The method has been extended to include a formula which has been derived to indicate the precision of the measured activities. The validity of this formula has been tested and found to be satisfactory. The techniques used to select appropriate counting conditions are described.
Similar content being viewed by others
References
J. S. ELDRIDGE, P. CROWTHER, Nucleonics, 22 (1964) 56.
D. L. HORROCKS, P. R. KLEIN, Nucl. Instr. Methods, 124 (1975) 585.
J. B. SCARBOROUGH, Numerical Mathematical Analysis, University Press, Oxford, 1950, p. 378.
D. L. HORROCKS, Clin. Chem., 21 (1975) 370.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Ediss, C., Abrams, D.N. Implementation of the coincidence method for determining125I activity with an estimate of error. J. Radioanal. Chem. 65, 341–347 (1981). https://doi.org/10.1007/BF02516117
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02516117