Abstract
Optically stimulated luminescence dosimeters (OSLDs) can be used as accurate and re-usable dosimeters for radiotherapy applications. OSLDs have been observed to decline in sensitivity with repeated use and it is important to determine whether this decline in sensitivity is associated with a decline in reliability. This study used three batches of OSLDs (purchased in 2012, 2014 and 2016) that had been repeatedly re-used in a mature in vivo dosimetry programme over a period of up to five years and evaluated the consistency of their response over repeated irradiation-readout-bleaching cycles. Each irradiation delivered 105 cGy to all OSLDs, using a 12 meV electron beam from a Varian iX linear accelerator. The five- and three-year-old OSLDs respectively displayed 86% and 89% of the sensitivity of the one year old OSLDs, but when a correction factor for each OSLD was derived based on the first measurement result and applied to each subsequent reading, all OSLDs were able to measure the 105 cGy test dose accurately, within standard deviations of 2.0% for the OSLDs from 2012 and 1.3% for the OSLDs from 2014 and 2016. If a mean calibration value was applied to the readings from each batch of OSLDs, instead of applying a measurement-derived sensitivity correction factor to each individual OSLD reading, the standard deviations increased to an unacceptable 6.1, 5.6 and 2.9%. Well-used three- and five-year-old OSLDs were shown to be capable of providing measurements with similar accuracy to a more recently-purchased batch of OSLDs, when measurement-derived sensitivity correction factors were applied to each result. If this extra step is included in the OSLD measurement process, then the same OSLDs may be reliably used for years without needing to be retired and replaced.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
Jursinic, P.A.: Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. Med. Phys. 34(12), 4594–4604 (2007). https://doi.org/10.1118/1.2804555.
Dunn, L., Lye, J., Kenny, J., et al.: Commissioning of optically stimulated luminescence dosimeters for use in radiotherapy. Radiat. Meas. 51, 31–39 (2013) https://doi.org/10.1016/j.radmeas.2013.01.012.
Villani, D., Mancini, A., Haddad, C.M., Campos, L.L.: Application of optically stimulated luminescence ‘nanoDot’ dosimeters for dose verification of VMAT treatment planning using an anthropomorphic stereotactic end-to-end verification phantom. Radiat. Meas. 106, 321–325 (2017). https://doi.org/10.1016/j.radmeas.2017.03.027.
Opp, D., Nelms, B.E., Zhang, G., Stevens, C. and Feygelman, V., 2013. Validation of measurement guided 3D VMAT dose reconstruction on a heterogeneous anthropomorphic phantom. Journal of Applied Clinical Medical Physics, 14(4), pp. 70–84. https://doi.org/10.1120/jacmp.v14i4.4154.
Charles, P.H., Crowe, S.B., Kairn, T., et al.: The effect of very small air gaps on small field dosimetry. Phys. Med. Biol. 57, 6947–6960 (2012). https://doi.org/10.1088/0031-9155/57/21/6947.
Scarboro, S.B., Followill, D.S., Kerns, J.R., et al.: Energy response of optically stimulated luminescent dosimeters for non-reference measurement locations in a 6 MV photon beam. Phys. Med. Biol. 57(9), 2505–2515 (2012). https://doi.org/10.1088/0031-9155/57/9/2505.
Peet, S.C., Wilks, R., Kairn, T., Crowe, S.B.: Measuring dose from radiotherapy treatments in the vicinity of a cardiac pacemaker. Phys. Medica. 32(12), 1529–1536 (2016). https://doi.org/10.1016/j.ejmp.2016.11.010.
Austin, M.J., Bergstrand, E.S., Bokulic, T., et al.: Development of Procedures for In VivoDosimetry in Radiotherapy. IAEA (International Atomic Energy Agency) Human Health Report 8 (2013).
Butson, M., Haque, M., Smith, L., et al.: Practical time considerations for optically stimulated luminescent dosimetry (OSLD) in total body irradiation. Australas. Phys. Eng. Sci. Med. 40(1), 167–171 (2017). https://doi.org/10.1007/s13246-016-0504-4.
Nabankema, S.K., Jafari, S.M., Peet, S.C., et al.: Wearable glass beads for in vivo dosimetry of total skin electron irradiation treatments. Radiat.Phys. Chem. 140, 314–318 (2017). https://doi.org/10.1016/j.radphyschem.2016.12.013.
Lye, J., Dunn, L., Kenny, J., et al.: Remote auditing of radiotherapy facilities using optically stimulated luminescence dosimeters. Med. Phys. 41(3), 032102 (2014). https://doi.org/10.1118/1.4865786.
Jursinic, P.A.: Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose. Med. Phys. 37(1), 132–140 (2010). https://doi.org/10.1118/1.3267489.
Asena, A., Crowe, S.B., Kairn, T., et al.: Response variation of optically stimulated luminescence dosimeters. Radiat. Meas. 61, 21–24 (2014). https://doi.org/10.1016/j.radmeas.2013.12.004.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Ethics declarations
The authors declare that they have no conflict of interest. This article does not contain any studies with human participants or animals performed by any of the authors.
Rights and permissions
Copyright information
© 2019 Springer Nature Singapore Pte Ltd.
About this paper
Cite this paper
Kairn, T., Peet, S., Yu, L., Crowe, S. (2019). Long-Term Reliability of Optically Stimulated Luminescence Dosimeters. In: Lhotska, L., Sukupova, L., Lacković, I., Ibbott, G. (eds) World Congress on Medical Physics and Biomedical Engineering 2018. IFMBE Proceedings, vol 68/3. Springer, Singapore. https://doi.org/10.1007/978-981-10-9023-3_103
Download citation
DOI: https://doi.org/10.1007/978-981-10-9023-3_103
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-10-9022-6
Online ISBN: 978-981-10-9023-3
eBook Packages: EngineeringEngineering (R0)